Dental Materials
Volume 24, Issue 8 , Pages 1095-1101, August 2008

Fatigue failure of as-received and retrieved NiTi orthodontic archwires

  • C. Bourauel

      Affiliations

    • Cendres+Métaux Endowed Chair of Oral Technology, School of Dentistry, Bonn University, Germany
  • ,
  • W. Scharold

      Affiliations

    • Department of Orthodontics, School of Dentistry, Bonn University, Germany
  • ,
  • A. Jäger

      Affiliations

    • Department of Orthodontics, School of Dentistry, Bonn University, Germany
  • ,
  • T. Eliades

      Affiliations

    • Department of Orthodontics, School of Dentistry, Aristotle University of Thessaloniki, Greece
    • Corresponding Author InformationCorresponding author. Department of Orthodontics, School of Dentistry, Aristotle University of Thessaloniki, 57 Agnoston Hiroon Street, Nea Ionia 14231, Greece.

Received 4 October 2007; received in revised form 17 December 2007; accepted 23 December 2007.

Abstract 

Aim

To investigate the fracture resistance of as-received and retrieved NiTi archwires.

Materials and methods

NiTi archwires (German Orthodontics, CA, USA) of various cross-sections were retrieved from orthodontic patients and brand-, type- and size-matched wires were included as controls. Specimens prepared from the selected wires were subjected to bending deformation at a loading frequency of 2Hz and cyclic loading was applied either until fatigue failure or with a maximum number of loading cycles of 2×106. For each wire cross-section a minimum of five specimens were loaded; the cycle number of the fatigue fracture was recorded and the mean of the cycles was calculated from the five specimens to provide the corresponding point in the Wöhler diagram. Scanning electron microscopy was used to assess the fracture pattern of broken wires. The results of the fracture cycles of as-received and retrieved wires were analyzed with a general linear regression model with state (as-received vs. retrieved) and wire size serving as discriminating variables and Tukey's test at the 0.05 level of significance.

Results

Retrieved wires fractured at a significantly lower number of cycles compared to their as-received matches. The size of the wire played a role in determining fracture with larger cross-sections showing reduced fatigue failure properties and 0.30 and 0.035-mm specimens showing no fracture at the selected strain and number of cycles. The SEM investigations revealed evidence of smoother fractured surfaces in large-sized, rectangular cross-sectioned retrieved wires. Kaplan–Meier survival analysis with log rank test demonstrated the effect of aging and size in predicting survival of NiTi wires.

Significance

The extended life expectancy of NiTi archwires proposed in current treatment trends is associated with higher probability of fatigue fracture of wires. Large diameter and square or rectangular cross-sections possess an increased chance of failure relative to smaller diameter wires. A necessity may emerge to monitor patients to identify wire failures despite the commonly employed practice of increased time intervals between appointments in patients treated with NiTi wires.

Keywords: Fatigue, Fracture, NiTi wires, Retrieval, Scanning electron microscopy

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 This paper was based on the Dr. Med. Dent. thesis of the late W. Scharold.

PII: S0109-5641(08)00004-3

doi:10.1016/j.dental.2007.12.007

Dental Materials
Volume 24, Issue 8 , Pages 1095-1101, August 2008